Product Detail

PrintPrint
Dynamic Flotation System

Huntleigh Nimbus 3 Advanced Dynamic Flotation System

An alternating air pressure reduction full mattress replacement. The mattress consists of deep, horizontal air cells without a side panel. The two way stretch cover is breathable and water resistant. An automatic electrically operated pump operates the air mattress. Available in adult single size and paediatric size. Designed to assist in the prevention and management of pressure ulcers for people at high to very high risk.

Learn More 
Features: Alternating Air, Water Resistant
  Dynamic Flotation System
Image Gallery: Dynamic Flotation System
Approximate Price: $14495
(3 suppliers)

More Details

The Nimbus 3 automatically adjusts to individual users regardless of their position, weight distribution and size. It works with infinite levels of adjustment.

STRUCTURE
This mattress replacement consists of figure of eight shaped air cells each of which run across the bed. Each cell is 200mm thick.
The mattress is made from heavy duty polyurethane and foam
The head section cells rare semi dynamic, while the other cells operate up and down in ten minute cycles.
Torso section adjusts so as to retain support without sinking in seated or lying positions.
It has a powered down heel section for improved pressure reduction.
Each cell can be removed individually for repairs and replacement.
It is designed for users at very high risk of pressure sore development and for use up to and including grade four pressure areas.

ATTACHMENT
Straps are used to secure the mattress replacement around the bed, with two additional straps at the top of the mattress.
Optional Divan Securing Straps for divan style single, double and king sized bed bases. This uses a buckle to secure.

PUMP AND CONTROLS
Separate to the mattress replacement with rear hooks to attach to bed end.
The pump alternately inflates and removes air pressure from the cells in a ten minute cycle.
The pressure is adjusted automatically using a sensor pad which monitors position, weight distribution and size of the user.
The comfort control allows manual adjustment of firm or soft settings.

COVER
Dartex cover has two way stretch, is vapour permeable (breathable), water resistant and loose fitting.
A zip on either side secures it into position.

SAFETY FEATURES
CPR facility situated on the top left side deflates the torso section in under ten seconds.
Transport mode duration of approximately twelve hours in static mode.
Audible and visual alarms indicate power on, loss of mains power, high air pressure and low air pressure. Battery back up for alarms and mute button.
A symbol appears when the Nimbus 3 needs to be serviced.
This mattress replacement has air filters.
Antifungal, antimicrobial and antibacterial agent is in the cover.
Protected zips and flat pump for cleaning.

CLEANING REQUIREMENTS
Wipe down clean in situ with warm soapy water.

SAFETY AND MAINTENANCE
Refer to the supplier and manufacturer manual for maintenance instructions and safety warnings.

CONSIDERATIONS FOR USE
Available in a paediatric or child model able to provide pressure reduction for those of small stature and low weight. The supplier states the paediatric mattress replacement has a minimum weight requirement of 6kg while the standard Nimbus has a minimum 25kg.

Factors To Consider

All pressure reduction products must be monitored to prevent 'bottoming out'. This occurs when the user's weight bears down so hard on the product material that it is flattened and the bony parts are pressed against the surface below the cushion or mattress. Where bottoming out occurs even with correct use (eg correct inflation levels), a different product may need to be considered.

Disclaimer

Whilst all care is taken to provide accurate information with respect to the item described, the Independent Living Centre (ILC) is not involved in product design or manufacture, and therefore not in a position to guarantee the accuracy of the information provided. Selection of equipment, which is both suitable and appropriate for individual needs remains the responsibility of the person(s) considering requisition, and no responsibility is taken by the ILC for any loss or injury caused through use of the equipment or alleged to have arisen through reliance upon information provided. As information is subject to change any enquiries should be directed to the manufacturer.

Suppliers

Supplier Location Contact Website
ArjoHuntleigh (NSW) Newington, NSW Sales: 1800 072 040
Fax: 02 9714 6333
Toll Free: 1800 072 040
  Visit Website 
ArjoHuntleigh (QLD) Virginia, QLD Toll Free: 1800 072 040
  Visit Website 
ArjoHuntleigh (SA) Stepney, SA Toll Free: 1800 072 040
  Visit Website 