Short Description
The Alpha Relief system is an alternating pressure relief system which consists of a pump and mattress replacement. Suitable for individuals at medium to high risk of developing pressure sores. The system comes with a carry bag. Available in single size only.
Options
Divan Securing Straps - tow straps with click buckle
Models
Only one model is available.
| Specification |
Mattress |
| Height |
133 mm |
| Length |
2090 mm |
| Load Capacity (supplier stated) |
160 kg |
| Weight (kg) |
Unknown |
| Width |
860 mm |
Additional Dimensions
Compressor
Weight (kg) NSW................................... 2.75 kg
Length NSW........................................ 248 mm
Height NSW........................................ 160 mm
Width NSW......................................... 248 mm
More Details
CONSTRUCTION
Cell material is polyurethane
17 fully alternating cells. Operating cycle is 10 minutes.
Cells are individually replaceable
Fits almost any hospital or domestic bed
COVER
Two way stretch, breathable (vapour permeable) cover which is water resistant and removable for cleaning.
Contains Actifresh - resists the development of mould and mildew and helps to prevent fabric degradation
Zips protected by a flap to prevent ingress of contaminants
ATTACHMENT
The mattress has securing straps for attachment to the bed.
A sheet can be loosely secured to the mattress with sheet clips to ensure that the sheet is tidy and safe.
PUMP
Lightweight and compact
Manual pressure setting control
Patient weight indicator for ease of use
Adjustable to patient weight
Adjustable to patient position
Minimal noise and vibration levels
Spring loaded hanging hooks
Integrated air filter
Specially designed snap lock connectors help prevent accidental disconnection of the mattress from the pump
Easy to use guide on pump
Slip resistant feet on base and back
Alarm indicators: audible and visual alarms, power on indicator, low pressure alarm, power fail alarm, inflation wait indicator, integral battery backup for alarm, self test for pump
SAFETY
CPR facility. The quick deflate pull is located at the top right hand sidde of the system. The mattress will deflate in 20 seconds.
Transport Facility allows the mattress to remain inflated for up to eight hours when not attached to the pump.
Factors To Consider
All pressure reduction products must be monitored to prevent 'bottoming out'. This occurs when the user's weight bears down so hard on the product material that it is flattened and the bony parts are pressed against the surface below the cushion or mattress. Where bottoming out occurs even with correct use (eg correct inflation levels), a different product may need to be considered.
Standards
Some assistive technology needs to meet Australian or other standards. Standards may relate to materials, manufacturing and installation. Products that meet Australian or international standards will have written certification. To find out if a product meets Australian Standards ask the supplier to show you the certificate. For more information about standards contact us or visit http://ilcaustralia.org.au/Using_Assistive_Technology/standards_regulation
Place Of Manufacture
UK